Product Search

#27700 GLP-1, Active form (High sensitivity) Assay Kit - IBL

  • correlation_between_#27784_#27700
  • correlation_between_#27700(without SPE)_Millipore (with SPE)
  • assay design_#27700_#27784
Intended Use:
Research reagents
Measuring Method:
ELISA
Sample Types:
Human, Mouse, Rat
Measuring Samples:
EDTA-plasma, Cell culture supernatant
Measurement Range:
0.25 ~ 16 pmol/L
Package Size1:
96 Well

※ The product indicated as "Research reagents" in the column Intended Use cannot be used
  for diagnostic nor any medical purpose.
※ The datasheet listed on this page is sample only. Please refer to the datasheet
  enclosed in the product purchased before use.

Product Overview

Product Overview

Product Code 27700
Product Name GLP-1, Active form (High sensitivity) Assay Kit - IBL
Maker Name Immuno-Biological Laboratories Co., Ltd.
Intended Use Research reagents
Measuring Method ELISA
Conjugate HRP
Species Human, Mouse, Rat
Measuring Samples EDTA-plasma, Cell culture supernatant
Measurement Range 0.25 ~ 16 pmol/L
Primary Reaction 2 ~ 8°C Overnight
Secondary Reaction 60 minutes at 2 ~8°C
Sensitivity 0.09 pmol/L
Specificity GLP-1 (7-36) amide 100%
GLP-1 (7-37)100%
GLP-1(1-37)0.04%
GLP-1(9-36) amide 0.02%
Glucagon N.D.
Human GLP-2 N.D.
Human GIP(1-42) N.D.
Human GIP(3-42) <0.01%
Mouse GIP(1-42) N.D.
Rat GIP(1-42) N.D.
Storage Condition 2~8°C
Measuring Service Available
Package Size 1 96 Well
Remarks1 *DPP4 inhibitor has to be added when collecting samples, or use purpose-made blood collection tubes in order to preserve GIP.

Product Description

Product Description

GLP-1 (Glucagon Like Peptide-1) is a transcriptional product of proglucagon gene as same as glucagon and one of gastrointestinal hormone which is secreted from L cells of lower section of the small intestine. GLP-1 (7-36) amide and GLP-1 (7-37) are active form and these peptides are dissolved in blood by DDP4 (dipeptidyl peptidase-4) into GLP-1 (9-36) amide and GLP-1 (9-37) and these peptides lose its activities in short period. GLP-1 is known as incretin as same as GIP (Glucose-dependent Insulinotropic Polypeptide), it promotes secretion of insulin from β cells of the isles of Langerhans in the pancreas as glucose dependently. GLP-1 also suppresses secretion of glucagon from α cell of pancreas and concentration of glucose in blood after meal. Drug of type 2 diabetes has been developed by utilizing the functional mechanism of GLP-1 such as blood sugar control through pancreas, delayed gastric emptying and appetite suppression which contributes weight loss as extra-pancreatic effects. This ELISA kit can quantitatively and specifically measure GLP-1, active form (GLP-1(7-36) amide and GLP-1 (7-37)).


Keywords:
GIP / GLP-1 / Insulin / Incretin

 

FAQ

FAQ

  • Question
    Q.What is the condition of DPP 4 inhibitor?
  • Answer
    A.Only EDTA-plasma collected using collecting tubes contained DPP4 inhibitor can be used for measuring of GLP-1 active form. BD™ P800 GLP-1, GIP, Glucagon, Ghrelin storing vacuumed collecting tube should be used for collecting blood or the blood should be blocked its activity by DPP4 inhibitor prior it is measured. The amount of DPP4 inhibitor for use is 10μL per 1mL blood or 20μL per 1mL plasma.
  • Question
    Q.Is composition of EIA buffer of each ELISA kit all same? Can it be mixed to use?
    ELISA common FAQ
  • Answer
    A.No it isn't. As constitute of each EIA buffer is different, it cannot be mixed with other lots or EIA buffers contained in other kind of ELISA kits.
  • Question
    Q.What is the composition of concentrated wash buffer?
    ELISA common FAQ
  • Answer
    A.It contains ordinary Tween and phosphate buffer (0.05% Tween-20 in PB).
  • Question
    Q.What is the feature of the plate?
    ELISA common FAQ
  • Answer
    A.We use plate that is flat bottom and removable strip type plate (8wellx 12 strips).
  • Question
    Q.Can I re-use standard after reconstitution?
    ELISA common FAQ
  • Answer
    A.Not recommended to re-use standard after reconstitution. Please use it at once after the reconstitution.
    Please note that there are some exceptions. One time freeze-thaw the standard is acceptable for use after reconstitution for some ELISAs.
    Please check the details on each product datasheet.
  • Question
    Q.What is different between reagent blank and test sample blank?
    ELISA common FAQ
  • Answer
    A.Reagent blank means a well is only added EIA buffer and the purpose is confirming whether the measuring value is influenced by lack of washing process. Test sample blank means a well is added EIA buffer and HRP antibody and the purpose is to calculate the background.
  • Question
    Q.How many samples can be measured by this kit?
    ELISA common FAQ
  • Answer
    A.The pre-coated plate contained in our ELISA kit is 96 wells plate. We recommend to use 16 wells (2 slits) for standard and 80 wells (10 slits) for 40 samples in duplicate.
  • Question
    Q.What is LOD (Limit of Detection)?
    ELISA common FAQ
  • Answer
    A.It (LOD) is defined as sensitivity that is calculated using the NCCSL method. Please refer to a datasheet of each product.
  • Question
    Q.What is LOQ (Limit of Quantification)?
    ELISA common FAQ
  • Answer
    A.It (LOQ) is the lowest value of measurement (standard) range. Please refer to a datasheet of each product.
  • Question
    Q.What is the definition of Over Night (O/N) reaction?
  • Answer
    A.It means that the reaction is required more than 16 hours unless otherwise specifically defined it on a datasheet of each ELISA product.