- HOME >
- For Researchers >
- Product Search >
- Search Result >
- #27261 Human DMP1 Assay Kit - IBL
Product Search
#27261 Human DMP1 Assay Kit - IBL
- Intended Use:
- Research reagents
- Measuring Method:
- ELISA
- Sample Types:
- Human
- Measuring Samples:
- Serum, EDTA-plasma, Urine, Cell culture supernatant
- Measurement Range:
- 3.6 ~ 230 pmol/L
- Package Size1:
- 96 Well
※ The product indicated as "Research reagents" in the column Intended Use cannot be used
for diagnostic nor any medical purpose.
※ The datasheet listed on this page is sample only. Please refer to the datasheet
enclosed in the product purchased before use.
Product Overview
Product Overview
Product Code | 27261 |
---|---|
Product Name | Human DMP1 Assay Kit - IBL |
Maker Name | Immuno-Biological Laboratories Co., Ltd. |
Intended Use | Research reagents |
Measuring Method | ELISA |
Conjugate | HRP |
Species | Human |
Measuring Samples | Serum, EDTA-plasma, Urine, Cell culture supernatant |
Measurement Range | 3.6 ~ 230 pmol/L |
Primary Reaction | Overnight at 2 - 8 ℃ |
Secondary Reaction | 60 minutes at 2 - 8 ℃ |
Sensitivity | 0.57 pmol/L |
Specificity | Substance Cross Reactivity Mouse DMP-1 ≦0.1% Rat DMP-1 ≦0.1% |
Storage Condition | 2 - 8 ℃ |
Poisonous and Deleterious Substances | Not Applicable |
Cartagena | Not Applicable |
Measuring Service | Available |
Package Size 1 | 96 Well |
Product Description
Product Description
Dentin Matrix Protein 1(DMP1) is a constitutive protein in extracellular matrix of osteocyte and it belongs to Small Integrin-Binding Ligand, N-linked Glycoprotein (SIBLING) family. It is synthesized as a full-length 105kDa proprotein and it is divided into 37kDa at N terminal side and 57kDa at C terminal side in the secretion process. It has a lot of acidic domains and it binds with calcium because of the feature that become negative charged within tissues and it has an important role for bone mineralization. DMP1 expresses in osteocyte while other constitutive proteins in extracellular matrix such as Osteopontin, Osteocalcin and Bone sialoprotein express in osteoblast cells. Osteocyte which is a main constituent cell of bone regulates activities of osteoclastic cells and osteoblast cells and it adjusts bone remodeling process. It also acts as an endocrine cell which secrets FGF-23. DMP1 in blood is considered as a candidate of biomarker which reflects activities of osteocyte and it is also suggested to become useful information for research of bone metabolism and phosphate metabolism. The concentration of DMP1 in human blood can be quantitatively measured by this ELISA kit.
FAQ
FAQ
-
Q.Is composition of EIA buffer of each ELISA kit all same? Can it be mixed to use?
ELISA common FAQ -
A.No it isn't. As constitute of each EIA buffer is different, it cannot be mixed with other lots or EIA buffers contained in other kind of ELISA kits.
-
Q.What is the composition of concentrated wash buffer?
ELISA common FAQ -
A.It contains ordinary Tween and phosphate buffer (0.05% Tween-20 in PB).
-
Q.What is the feature of the plate?
ELISA common FAQ -
A.We use plate that is flat bottom and removable strip type plate (8wellx 12 strips).
-
Q.Can I re-use standard after reconstitution?
ELISA common FAQ -
A.Not recommended to re-use standard after reconstitution. Please use it at once after the reconstitution.
Please note that there are some exceptions. One time freeze-thaw the standard is acceptable for use after reconstitution for some ELISAs.
Please check the details on each product datasheet. -
Q.What is different between reagent blank and test sample blank?
ELISA common FAQ -
A.Reagent blank means a well is only added EIA buffer and the purpose is confirming whether the Test sample value is influenced by lack of washing process or other operations. Test sample blank means a well is added EIA buffer and HRP antibody and the purpose is to calculate the background.
-
Q.How many samples can be measured by this kit?
ELISA common FAQ -
A.The pre-coated plate contained in our ELISA kit is 96 wells plate. We recommend to use 16 wells (2 slits) for standard and 80 wells (10 slits) for 40 samples in duplicate.
-
Q.What is LOD (Limit of Detection)?
ELISA common FAQ -
A.It (LOD) is defined as sensitivity that is calculated using the NCCSL method. Please refer to a datasheet of each product.
-
Q.What is LOQ (Limit of Quantification)?
ELISA common FAQ -
A.It (LOQ) is the lowest value of measurement (standard) range. Please refer to a datasheet of each product.
-
Q.What is the definition of Over Night (O/N) reaction?
ELISA common FAQ -
A.It means that the reaction is required more than 16 hours unless otherwise specifically defined it on a datasheet of each ELISA product.
-
Q.What is the specification of quality control for ELISA product release?
ELISA common FAQ -
A.The information of specification is available on individual lot specific CoA. Please contact us with your reference lot number for obtaining of specific CoA.
-
Q.What is the number (e.g. 432143214321) at the edge of strips of the plate?
ELISA common FAQ -
A.According to the plate maker (ThermoFisher), it does not have any specific meaning as it is just the number of molds.
-
Q.How to wash an ELISA plate?
ELISA common FAQ -
A.Washing it by an auto-washer is highly recommended.
If it is not available, please refer to the demo video (only 2 mins) using a washing bottle. -
Q.The wells turned black during the test with the kit.
ELISA common FAQ -
A.It is possible that the wells were not washed sufficiently during the washing process after the HRP-labeled antibody reaction.
Be sure to wash the wells enough times as described in the data sheet with washing buffer of more than 350 µL.