ELISA - IBL()
-
Q.Can it be measured in urine sample?
-
A.Yes, it is measurable, but it is unstable.
The dilution ratio should be evaluated and determined by each laboratory, but it is considered as it might be in the range between 10,000 and 20,000 fold. In this case, the commercially available product specific dilution buffer (Code: 27799D-30mL) is additionally required.
uired.
-
Q.What is the expression of the standard?
-
A.It is a recombinant protein expression, which used CHO cells.
-
Q.What is the recommended dilution factor for serum?
-
A.The suitable dilution factor may vary by the condition, but we generally use around x5,000 for the measurements.
-
Q.What are the epitopes of the plate-coated antibody used in this kit?
-
A.Antibody is from recombinant protein antigen. No epitope mapping has conducted.
-
Q.What is the epitope of the labeled antibody?
-
A.Antibody is from recombinant protein antigen. No epitope mapping has conducted.
-
Q.Are normal controls measurement data and reproducibility data available?
-
A.Yes, we have. The date is available on the following product page. Please refer to the Pamphlet 2 listed on the page.
http://www.ibl-japan.co.jp/en/search/product/detail/id=5671 -
Q.Does this kit specifically detect LRG1 or other LRG families as well?
-
A.We have not checked the cross reactivity against other LRG families using our LRG-1 kit (#27798).
-
Q.Does this kit cross react with mouse?
-
A.Yes, it does.
-
Q.Does this kit cross react with mouse?
-
A.No, it does not.
-
Q.Does this kit cross react with mouse?
-
A.Yes, it does.
-
Q.Can mouse secreted α-Klotho be measured by this ELISA kit?
-
A.No, it cannot.
-
Q.Can secreted α-Klotho in supernatant can be detected by this kit?
-
A.Yes, it can be detected.
-
Q.What is the difference between soluble and secreted α-Klotho kits?
-
A.The α-Klotho protein has been reported to shedding its long N-terminal extracellular domain, which occupies most of the sequence, and to be released in the blood (soluble). At the same time, gene splicing has been reported to produce a secreted isoform with only the first repeat (secreted).
#27998 Human soluble α-Klotho Assay Kit - IBL specifically detects all soluble α-Klotho including the KL1 domain.
#27901 Human secreted α-Klotho Assay Kit - IBL specifically detects secreted α-Klotho in the KL1 domain, including the sequence QVSQLTKPISSLTKPYH.
-
Q.What is the immunogen used for the antibodies?
-
A.It is full length recombinant ApoA5.
-
Q.What is the origin of the standard enclosed in the kit?
-
A.It is purified recombinant protein derived from CHO.
It does not contain human-derived substances (blood, tissue, etc.).
- 99999
- 99999-1 Control Set - IBL common FAQ
-
Q.How do you define the acceptance range of quality control of control sets?
Control Set Common FAQ -
A.In principal, it is defined as ±15%-20% based the standard value described on each datasheet.
-
Q.What is the shelf life of control set?
Control Set Common FAQ -
A.The shelf life is 12 months upon the goods is dispatched from our lab if the goods is unopned.
-
Q.What is the storage condition of control sets?
Control Set Common FAQ -
A.It should be stored under chilled condition (2~10℃).
-
Q.Can the control set be used after reconstitution?
Control Set Common FAQ -
A.No it cannot be re-used after the reconstitution.
-
Q.What is the package form of control sets?
Control Set Common FAQ -
A.It is lyophilized form.
-
Q.How should I conduct Endothelin extraction process?
-
A.Please refer to datasheet enclosed the product package and/or the datasheet listed on the product page.
As several publications can be found on the page (product description), please refer to them. -
Q.Is extraction process essential for blood samples?
-
A.Yes, it is essential. Please follow the datasheet for the extraction process using column such as SepPak.
-
Q.I have only 0.5mL blood samples. Is it possible to measure it using this ELISA?
-
A.Yes, it is possible, however, in general speaking, the concentration of Endothelin in blood sample is low conc., the concentration could be much lower after the extraction and process. Please well consider the risk of lack of concentration prior start the measurement.
Our recommendation is more than 2mL blood samples.
-
Q.Can urine samples be used for this kit?
-
A.No, it cannot.
-
Q.Can urine samples be used for this kit?
-
A.No, it cannot.
-
Q.Can urine samples be used for this kit?
-
A.No, it cannot.
-
Q.Can urine samples be used for this kit?
-
A.No, it cannot.
-
Q.The specificity on the datasheet is described as "Endothelin-1 is ≦0.1%".
What does exactly mean of "Endothelin-1" described here? -
A.It means Endothelin-1(1-21).
-
Q.How should I conduct Endothelin extraction process?
-
A.Please refer to the datasheet enclosed the product package and/or the datasheet listed on the product page.
As several publications can be found on the page (product description), please refer to them.
https://www.ibl-japan.co.jp/en/search/product/detail/id=3873 -
Q.Is extraction process essential for blood samples?
-
A.Yes, it is essential. Please follow the datasheet for the extraction process using column such as SepPak.
-
Q.I have only 0.5mL blood samples. Is it possible to measure it using this ELISA?
-
A.Yes, it is possible, however, in general speaking, the concentration of Endothelin in blood sample is low conc., the concentration could be much lower after the extraction and process. Please well consider the risk of lack of concentration prior start the measurement.
Our recommendation is more than 2mL blood samples.
-
Q.Which domain is recognized?
-
A.It is LCCL domain.
-
Q.If urine samples are not measured immediately after collection, how should they be stored?
-
A.They should be store at -20°C or below immediately after collection.
-
Q.Is any normalization needed for measured value of urine sample?
-
A.Yes, it is needed.
Please measure urinary creatinine at the same time and normalize to it. -
Q.Can rodent (mouse or rat) sample be used for this kit?
-
A.No, it cannot.
-
Q.How many time freeze-thaw can be repeated for human urine sample for the assay?
-
A.It is stable up to 3 times freeze-thaw repeat. Please refer to the freeze-thaw stability test result from this link.
-
Q.Do you have any storage (temperature) stability data of urinary titin N-terminal fragment (UTF)?
-
A.Yes, we do. Please refer to the data.
-
Q.Is GROβ same as CXCL2?
-
A.Yes, it is.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL). -
Q.What peptide is used for Standard enclosed in the kit (27418)?
-
A.It's synthetic peptide N3pE 3-40.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL). -
Q.What peptide is used for Standard enclosed in the kit (27716)?
-
A.It's synthetic peptide N3pE 3-42.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL).
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL).
-
Q.What is the epitope of antibodies used in this kit?
-
A.Anti-Human Mac-2bp (8A2) Mouse IgG (Precoated Antibody) recognizes position no. 182-276 amino acid of human Mac-2bp.
Anti- Human Mac-2bp (67A1) Mouse IgG (HRP Antibody) recognizes position no. 361-440 amino acid of human Mac-bp. -
Q.What is the immunogen of 2 antibodies used in the kit?
-
A.Recombinant protein was used as the immunogen for both antibodies.
-
Q.Can this kit be used for mouse or rat samples?
-
A.No, it cannot.
A monoclonal antibody (82E1) which is used as the labeled antibody in this kit, specifically reacts to N-terminal of Human Aβ.
Since the N-terminal sequence of Aβ is different between human and rodents (mouse or rat), it does not cross reacts to mouse and/or rat Aβ.
-
Q.Can this kit be used for bovine samples?
-
A.No, it cannot.
-
Q.Can this kit be used for bovine samples?
-
A.No, it cannot.
-
Q.Does this kit detect Insulin I or Insulin II?
-
A.This kit can be used for quantitatively measuring total insulin including Insulin I, Insulin II and Proinsulin.
-
Q.Does hemolyzed sample affect to the assay result?
-
A.Yes it does. We recommend to avoid to use hemolyzed samples for the assay as it might become value due to the effect of insulin-degrading enzymes leaked from blood cells.
-
Q.What is used for the STD?
-
A.Human IgA1 is treated with neuraminidase and galactosidase and used as Gd-IgA1.
Please refer to the following reference.
Novel lectin-independent approach to detect galactose-deficient IgA1 in IgA nephropathy. Yasutake J et al. Nephrol Dial Transplant. 2015 Aug;30(8):1315-21. https://pubmed.ncbi.nlm.nih.gov/26109484/ -
Q.Does Neuramidase treatment require for the sample?
-
A.No it doesn't. There is no much different for the result between with and without Neuramidase.
-
Q.I found the recommended dilution ratio is between 200 and 800 fold. That is quite a wide range. What is the most optimized dilution ratio?
-
A.Some samples have a very high value of Gd-IgA1 and such samples would be range over that why the recommended dilution ratio is wide.
We recommend to start from 200 fold dilution and find the best dilution ratio for your samples. The optimized dilution ratio should be determined by each user. -
Q.What is approx. normal IgA concentration in blood?
-
A.Total IgA is approx 1-4mg/mL in blood.
-
Q.What is approx. normal Gd-IgA1 concentration in blood?
-
A.As it is approx 1,000 ng/mL in blood, more than 200 fold dilution is recommended for measurement.
-
Q.What is the dilution ratio for urine sample?
-
A.Recommended dilution ratio is 2-4 fold using “4 EIA buffer” enclosed in this kit.
-
Q.How the calibration curve is made?
-
A.The calibration curve is made by detecting only Gd-IgA1 from the standard.
-
Q.Is control set of this kit available?
-
A.Unfortunately, it is not available due to raw material nature.
-
Q.What is epitope of the antibodies used in this kit?
-
A.The HRP conjugated labeled antibody (Anti-Human APP (R101A4) Mouse IgG Fab’ Affinity Purify) used in this kit detects domain at N-Terminal.
The epitope of the precoated plate antibody (Anti-Human sAPPα (2B3) Mouse IgG MoAb Affinity Purify at C-Terminal) used in this kit is DAEFRHDSGYEVHHQK. -
Q.Does this kit cross reacts to mouse sAPPα if TG mice that express human sAPPα is used for the assay?
-
A.No, it does not cross react to mouse sAPPα. This kit only detects human sAPPα.
-
Q.What the standards are derived from?
-
A.It is recombinant protein originated from CHO.
-
Q.What is the optimal dilution ratio for CSF samples measurement?
-
A.Please refer to the section "PERFORMANCE CHARACTERISTICS" on page 2 of the datasheet.
2-4 fold dilution is recommended to start for finding the optimal dilution ratio.
https://www.ibl-japan.co.jp/files/topics/1670_ext_02_en_0.pdf -
Q.What is the optimal dilution ratio for EDTA-plasma samples measurement?
-
A.Please refer to the section "PERFORMANCE CHARACTERISTICS" on page 2 of the datasheet.
2-4 fold dilution is recommended to start for finding the optimal dilution ratio.
https://www.ibl-japan.co.jp/files/topics/1670_ext_02_en_0.pdf -
Q.What is the optimal dilution ratio for Serum samples measurement?
-
A.Please refer to the section "PERFORMANCE CHARACTERISTICS" on page 2 of the datasheet.
2-4 fold dilution is recommended to start for finding the optimal dilution ratio.
https://www.ibl-japan.co.jp/files/topics/1670_ext_02_en_0.pdf
- 27752
- BACE1 Assay Kit - IBL
-
Q.Does this kit cross react to BACE2?
-
A.Not, it doesn't.
-
Q.What is recommended dilution ratio for human serum samples?
-
A.The recommended dilution ratio for human serum sample is 2 fold.
-
Q.What type of soluble APP can be measured by this kit?
-
A.This kit cross reacts to any type of human soluble APP.
-
Q.Does it cross react to mouse or rat Aβ (1-42)?
-
A.No it doesn't.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL).
-
Q.Does it cross react to mouse or rat Aβ (1-40)?
-
A.No it doesn't.
-
Q.Can frozen serum sample be used for this kit?
-
A.Yes, it can. However, stability test of the samples should be conducted and confirmed by each institute.
-
Q.How should I conduct Endothelin extraction process?
-
A.Please refer to the datasheet enclosed the product package and/or the datasheet listed on the product page.
As several publications can be found on the page (product description), please refer to them.
https://www.ibl-japan.co.jp/en/search/product/detail/id=3868 -
Q.Is extraction process essential for blood samples?
-
A.Yes, it is essential. Please follow the datasheet for the extraction process using column such as SepPak.
-
Q.I have only 0.5mL blood samples. Is it possible to measure it using this ELISA?
-
A.Yes, it is possible, however, in general speaking, the concentration of Endothelin in blood sample is low conc., the concentration could be much lower after the extraction and process. Please well consider the risk of lack of concentration prior start the measurement.
Our recommendation is more than 2mL blood samples.
-
Q.What is the condition of DPP 4 inhibitor?
-
A.Only EDTA-plasma collected using collecting tubes contained DPP4 inhibitor can be used for measuring of GLP-1 active form. BD™ P800 GLP-1, GIP, Glucagon, Ghrelin storing vacuumed collecting tube should be used for collecting blood or the blood should be blocked its activity by DPP4 inhibitor prior it is measured. The amount of DPP4 inhibitor for use is 10μL per 1mL blood or 20μL per 1mL plasma.
-
Q.This kit can be used for measurement of diabetes drug (e.g. Liragrutide)?
-
A.As this kit was developed to measure native GLP-1 active form (7-36/37), we have not evaluated the kit with any commercially available drugs.
Please contact us if you wish to evaluate the kit for the purpose.
-
Q.What is recommended dilution ratio for serum samples?
-
A.It is 2 to 8 fold.
-
Q.Can this kit specifically measure phosphorylated α-Synuclein?
-
A.No, this kit can measure any type of α-Synuclein. It is not specific for phsphorylated form.
-
Q.Can mouse or rat samples be used for this kit?
-
A.Wild type mouse or rat samples cannot be used, but APP-transgenic (Tg2576) mouse can be used for the kit.
-
Q.Is there any recommended buffer solution for cell extraction?
-
A.We recommend to use our product #19022 IBLysis.
-
Q.What is the epitope of the antibodies used in this kit?
-
A.The epitope of HRP conjugated labeled antibody (Anti-Human Aβ (N) (82E1) Mouse IgG MoAb Affinity Purify at N-Terminal) used in this kit is DAEFRHDSGYEVHHQK.
The epitope of the precoated plate antibody (Anti-Human APP-C Rabbit IgG Affinity Purify at C terminal) used in this kit is NGYENPTYKFFEQMQN.
-
Q.Can serum samples be used for this kit?
-
A.No, it cannot.
-
Q.Can serum samples be used for this kit?
-
A.Not recommended to use serum samples. Please use EDTA-plasma samples.
-
Q.What is recommended anticoagulant?
-
A.We use 2Na.
-
Q.Can VEGF-R2 be measured by this kit?
-
A.No, it cannot.
-
Q.Can serum sample be used for this kit?
-
A.Not recommended to use serum samples. Please refer to the following article for comparison data between serum and plasma. https://www.ncbi.nlm.nih.gov/pubmed/?term=25045139
-
Q.What is recommended dilution ratio?
-
A.8 fold dilution is acceptable for the measurement, however, we recommend to dilute mouse urine sample more than 16 fold to avoid the data becomes range over.
-
Q.Can rat sample be used for this kit?
-
A.No, it cannot.
-
Q.How mouse urine sample can be stored?
-
A.It should be stored at -80℃ immediately after collection.
-
Q.Is stability data available?
-
A.It is available on the following relevant product page as images.
http://www.ibl-japan.co.jp/en/search/product/detail/id=4678 -
Q.Can this kit measure mouse blood sample?
-
A.No, it cannot.
-
Q.How can we collect mouse urine sample?
-
A.The method of collection of mouse urine is listed on the relevant product page.
Please refer to the Pamphlet 2 listed on the page.
http://www.ibl-japan.co.jp/en/search/product/detail/id=4678 -
Q.What are the epitopes of pre-coated antibody (Anti-mouse Klotho 162A1 Rat IgG) and HRP conjugated antibody (Anti-mouse Klotho 235A1 Rat IgG Fab')?
-
A.Klotho has two domains (KL1 and KL2). Although we have not conducted epitope mapping, these antibodies can detect KL1 domains.
The kit was tested using culture supernatants expressing only KL1 of mouse Klotho and has shown a positive response. -
Q.Can dog sample (blood or urine) be used for this kit?
-
A.No it cannot.
-
Q.Which isoform of FGF21 can detect by this kit?
-
A.This assay kit react to both full length FGF 21 (1-181) and FGF21 (7-181). Therefore, this kit can measure total FGF21.
-
Q.How much is the volume of sample required for measurement?
-
A.5μL is recommended.
-
Q.How long does the plate coated biotinylated antibody become stable for use?
-
A.We have confirmed that it is stable for a week.
-
Q.How long does streptavidin plate become stable for use after first opening the pack?
-
A.We have confirmed that it is stable at 37℃ for 2 weeks.
-
Q.How much is the volume of sample required for measurement?
-
A.5μL is recommended.
-
Q.How long does the plate coated biotinylated antibody become stable for use?
-
A.We have confirmed that it is stable for a week.
-
Q.How long does streptavidin plate become stable for use after first opening the pack?
-
A.We have confirmed that it is stable at 37℃ for 2 weeks.
-
Q.How much is the volume of sample required for measurement?
-
A.2μL is acceptable for measuring GIP Total using samples collected after meal, however at least 5μL is required for measuring it using samples collected in fasting.
-
Q.How long does the plate coated biotinylated antibody become stable for use?
-
A.We have confirmed that it is stable for a week.
-
Q.How long does streptavidin plate become stable for use after first opening the pack?
-
A.We have confirmed that it is stable at 37℃ for 2 weeks.
-
Q.How much is the volume of sample required for measurement?
-
A.2μL is acceptable for measuring GIP Total using samples collected after meal, however at least 5μL is required for measuring it using samples collected in fasting.
-
Q.How long does the plate coated biotinylated antibody become stable for use?
-
A.We have confirmed that it is stable for a week.
-
Q.How long does streptavidin plate become stable for use after first opening the pack?
-
A.We have confirmed that it is stable at 37℃ for 2 weeks.
- 99999
- 99999 ELISA - IBL common FAQ
-
Q.Is composition of EIA buffer of each ELISA kit all same? Can it be mixed to use?
ELISA common FAQ -
A.No it isn't. As constitute of each EIA buffer is different, it cannot be mixed with other lots or EIA buffers contained in other kind of ELISA kits.
-
Q.What is the composition of concentrated wash buffer?
ELISA common FAQ -
A.It contains ordinary Tween and phosphate buffer (0.05% Tween-20 in PB).
-
Q.What is the feature of the plate?
ELISA common FAQ -
A.We use plate that is flat bottom and removable strip type plate (8wellx 12 strips).
-
Q.Can I re-use standard after reconstitution?
ELISA common FAQ -
A.Not recommended to re-use standard after reconstitution. Please use it at once after the reconstitution.
Please note that there are some exceptions. One time freeze-thaw the standard is acceptable for use after reconstitution for some ELISAs.
Please check the details on each product datasheet. -
Q.What is different between reagent blank and test sample blank?
ELISA common FAQ -
A.Reagent blank means a well is only added EIA buffer and the purpose is confirming whether the Test sample value is influenced by lack of washing process or other operations. Test sample blank means a well is added EIA buffer and HRP antibody and the purpose is to calculate the background.
-
Q.How many samples can be measured by this kit?
ELISA common FAQ -
A.The pre-coated plate contained in our ELISA kit is 96 wells plate. We recommend to use 16 wells (2 slits) for standard and 80 wells (10 slits) for 40 samples in duplicate.
-
Q.What is LOD (Limit of Detection)?
ELISA common FAQ -
A.It (LOD) is defined as sensitivity that is calculated using the NCCSL method. Please refer to a datasheet of each product.
-
Q.What is LOQ (Limit of Quantification)?
ELISA common FAQ -
A.It (LOQ) is the lowest value of measurement (standard) range. Please refer to a datasheet of each product.
-
Q.What is the definition of Over Night (O/N) reaction?
ELISA common FAQ -
A.It means that the reaction is required more than 16 hours unless otherwise specifically defined it on a datasheet of each ELISA product.
-
Q.What is the specification of quality control for ELISA product release?
ELISA common FAQ -
A.The information of specification is available on individual lot specific CoA. Please contact us with your reference lot number for obtaining of specific CoA.
-
Q.What is the number (e.g. 432143214321) at the edge of strips of the plate?
ELISA common FAQ -
A.According to the plate maker (ThermoFisher), it does not have any specific meaning as it is just the number of molds.
-
Q.How to wash an ELISA plate?
ELISA common FAQ -
A.Washing it by an auto-washer is highly recommended.
If it is not available, please refer to the demo video (only 2 mins) using a washing bottle. -
Q.The wells turned black during the test with the kit.
ELISA common FAQ -
A.It is possible that the wells were not washed sufficiently during the washing process after the HRP-labeled antibody reaction.
Be sure to wash the wells enough times as described in the data sheet with washing buffer of more than 350 µL.
-
Q.What is the condition of DPP 4 inhibitor?
-
A.Only EDTA-plasma collected using collecting tubes contained DPP4 inhibitor can be used for measuring of GLP-1 active form. BD™ P800 GLP-1, GIP, Glucagon, Ghrelin storing vacuumed collecting tube should be used for collecting blood or the blood should be blocked its activity by DPP4 inhibitor prior it is measured. The amount of DPP4 inhibitor for use is 10μL per 1mL blood or 20μL per 1mL plasma.
-
Q.What is recommended dilution ratio for blood samples?
-
A.Please start to dilute from 2 fold.
-
Q.What is recommended dilution ratio for human samples of healthy subject?
-
A.It is 2 to 4 fold.
-
Q.Can this kit measure any drug for type 2 diabetes such as Liraglutide?
-
A.Liraglutide can be measured by this kit, however any other drugs are unknown.
-
Q.How can I store the standard after reconstitute it?
-
A.It cannot be recommended to be stored after reconstitution as it is designed for use one-off. However, if you need to reuse it due to uncontrolable situation, please store it in frozen.
-
Q.What size of oligomer can be detected by this assay kit?
-
A.Any oligomer that is bigger than Amyloidβ dimer and has a recognized region of 82E1 (N-terminal of 1-16) can be detected by this assay kit.
-
Q.Can this kit detect protofibril and fibril?
-
A.It cross reacts with both protofibril and fibril.
-
Q.3 recognized parts of 82E1 antibody (a.a. 1-16) are existed if the sample is trimer. In that case, does the conjugated antibody binds to 2parts or only one part?
-
A.82E1 antibody is used for as capture antibody (precoated one) and detection antibody (HRP conjugated one). Theoretically, 2 regions of trimer can bind to the detection antibody if 1 region is captured by captured antibody. However, this is only the case that actual molecule detected holds the N-terminal region of Aβ1-16.
-
Q.Is it depending on the size of oligomer to determine the quantity of the detection antibody that binds to one oligomer?
-
A.It depends on molecule of oligomer. It does not detect monomer. We recommend to use both products #27718 Human Amyloidβ (1-40) (FL) Assay Kit - IBL and #27719 Human Amyloidβ (1-42) (FL) Assay Kit - IBL by depending on purposes.
-
Q.What is recommended dilution ratio for serum samples?
-
A.It is 2 to 8 fold.
-
Q.Can you verify that it is HIV or HBV negative?
-
A.The plasma used as the material was verified as negative and not cross reactive against HbsAG, anti-HCV, anti-HBc by the method of approved by FDA.
-
Q.What is used for the standard?
-
A.It is ApoB 100 that is purified from human plasma.
-
Q.Does this assay kit detect any specific type of VEGF or any type of VEGF such as A, B, C and D?
-
A.It cross reacts with VEGF-A. Isoforms like 120, 164 and 188 can be measured as mixed substances if the sample is lysate.
-
Q.Is it possible to measure VEGF expressed in lysate of mice muscle?
-
A.If it is an ordinary tissue lysate, it can be measured by the assay kit.
-
Q.What is the structure and host of the standard?
-
A.It is originated from CHO supernatant and it is full length Flag-Tag.
-
Q.Can this kit be used for serum?
-
A.No it cannot be used for normal serum.
-
Q.Why extraction and condensation process using Sep-Pack is required for measuring Endothelin-3 (ET-3) blood samples?
Is the process essential for the assay? -
A.Measuring ET-3 in blood sample is quite difficult due to very low concentration of ET-3 (reference approx. 0.48pg/mL in normal subject blood sample).
The extraction and condensation process using Sep-Pack is essential for the assay.
The assay will be failed without the process.
The ET-3 in blood sample become much lower than the original concentration after the process and it depends on the collection rate.
Therefore, it is recommended to prepare larger volume 5mL-10mL of blood sample for ET-3 measurement.
Please refer to "Attention for sample handling" on page 2 of the datasheet for additional information. -
Q.Which region are recognized by the antibodies used by this kit?
-
A.The capture antibody (Anti-ET 15-21) recognizes the region at C-terminal of ET-3 and the detection antibody specifically recognizes the region at N-terminal of ET-3.
-
Q.How should I conduct Endothelin extraction process?
-
A.Please refer to the datasheet enclosed the product package and/or the datasheet listed on the product page.
As several publications can be found on the page (product description), please refer to them.
https://www.ibl-japan.co.jp/en/search/product/detail/id=3869 -
Q.Is extraction process essential for blood samples?
-
A.Yes, it is essential. Please follow the datasheet for the extraction process using column such as SepPak.
-
Q.I have only 0.5mL blood samples. Is it possible to measure it using this ELISA?
-
A.Yes, it is possible, however, in general speaking, the concentration of Endothelin in blood sample is low conc., the concentration could be much lower after the extraction and process. Please well consider the risk of lack of concentration prior start the measurement.
Our recommendation is more than 2mL blood samples.
-
Q.Can this kit measure the full length of α-Klotho?
-
A.Yes, it can be used to measure the full length α-Klotho and KL1.
-
Q.What are the epitopes of pre-coated antibody (Anti-human Klotho (67G3) Mouse IgG MoAb A.P.) and HRP conjugated antibody (Human Klotho (91F1) Mouse IgG MoAb Fab’ A.P.)?
-
A.Klotho has two domains (KL1 and KL2). Although we have not conducted epitope mapping, these antibodies can detect KL1 domains.
-
Q.Do you have any information regarding the specificity of the kit?
-
A.Please refer to Fig 1 on page 10 of the following article (PMID:20599764).
Establishment of sandwich ELISA for soluble alpha-Klotho measurement: Age-dependent change of soluble alpha-Klotho levels in healthy subjects. Yamazaki Y et al. Biochem Biophys Res Commun. 2010 Jul 30;398(3):513-8. -
Q.Does this kit cross react with β-Klotho?
-
A.No, it does not. As the data is listed on the product page, please refer to it.
https://www.ibl-japan.co.jp/en/search/product/detail/id=3959 -
Q.Can this ELISA detect other species rather than human such as canine (dog), rabbit, rat and mouse?
-
A.No it cannot. However, cynomolgus samples is possible to be measured. Mouse urine samples can be measured by #27601 Mouse soluble α-Klotho Assay Kit - IBL.
-
Q.What kind of plasma sample can be used?
-
A.Typical data of the performance of the kit were obtained using human serum, EDTA-plasma, urine and cell culture supernatant for this kit, however, as you can see the data of heparin-plasma and citrated plasma measured by this kit displayed on this product page, they can be also measured by this kit.
-
Q.Is any sample stability data available?
-
A.Yes, it is available on the following relevant product page. Please refer to the Pamphlet listed on the page.
http://www.ibl-japan.co.jp/en/search/product/detail/id=3959 -
Q.What is pM unit of standards?
-
A.The 6,000pg/mL of standard in this kit corresponds to 55pM (as Molecular Weight 108.99kDa). Please refer to the article Establishment of sandwich ELISA for soluble alpha-Klothomeasurement: Age-dependent change of soluble alpha-Klotho levels in healthy subjects YYamazaki et al. Biochem Biophys Res Commun. 2010 July 30; 398(3): 513–518 (PMID 20599764 / https://www.ncbi.nlm.nih.gov/pubmed/?term=20599764 ) for further information of development of this kit.
-
Q.What is the difference between soluble and secreted α-Klotho kits?
-
A.The α-Klotho protein has been reported to shedding its long N-terminal extracellular domain, which occupies most of the sequence, and to be released in the blood (soluble). At the same time, gene splicing has been reported to produce a secreted isoform with only the first repeat (secreted).
#27998 Human soluble α-Klotho Assay Kit - IBL specifically detects all soluble α-Klotho including the KL1 domain.
#27901 Human secreted α-Klotho Assay Kit - IBL specifically detects secreted α-Klotho in the KL1 domain, including the sequence QVSQLTKPISSLTKPYH.
-
Q.Can this kit be used for mouse or rat samples?
-
A.No, they cannot.
-
Q.What is the epitope of antibodies used in this ELISA?
-
A.Since a whole recombinant human G-CSF was used as the immunogen for these antibodies and no epitope mapping has been done, the information is not available.
-
Q.Can this kit detect enzymatic activity?
-
A.No it cannot.
-
Q.Can this kit use for cow, mouse or rat samples?
-
A.No, it cannot.
-
Q.What is epitope of the antibodies used in this kit?
-
A.The epitope of the HRP conjugated labeled antibody (Anti-Human APP (R101A4) Mouse IgG Fab’ Affinity Purify) used in this kit has not been confirmed.
The epitope of the precoated plate antibody (Anti-Human sAPPβ-wild Type Rabbit IgG Affinity Purify at C-Terminal) used in this kit is ISEVKM. -
Q.What type of soluble APP can be measured by this kit?
-
A.This kit specifically detects human soluble APPβ wild type (sAPPβ-w).
-
Q.What the standards are derived from?
-
A.It is recombinant protein originated from CHO.
-
Q.What is the optimal dilution ratio for CSF samples measurement?
-
A.Please refer to the section "PERFORMANCE CHARACTERISTICS" on page 2 of the datasheet.
2-4 fold dilution is recommended to start for finding the optimal dilution ratio.
https://www.ibl-japan.co.jp/files/topics/1668_ext_02_en_0.pdf -
Q.What is the optimal dilution ratio for EDTA-plasma samples measurement?
-
A.Please refer to the section "PERFORMANCE CHARACTERISTICS" on page 2 of the datasheet.
2-4 fold dilution is recommended to start for finding the optimal dilution ratio.
https://www.ibl-japan.co.jp/files/topics/1668_ext_02_en_0.pdf -
Q.Can this kit measure sAPPβ in serum samples?
-
A.No, it is not recommended. sAPPβ in serum cannot be measured by this kit. Please use EDTA-plasma samples for the kit.
-
Q.Can this kit be used for heparin-plasma samples?
-
A.Endothelin-1 in heparin plasma can be measured. However, the extraction process must be performed. For details about the extraction process, see “12. Endothelin extraction process” on page 2 of the datasheet. Please refer to the datasheet for additional information.
-
Q.What kind of animal samples can be used with this kit?
-
A.Human, rat, pig, canine, mouse, cow, rabbit samples can be used with the kit as epitope of No.1 to 21 of Endothelin-1 is same among the species.
-
Q.In regards to the extraction of the plasma sample, can I use a column other than C18?
-
A.Please ask the maker for the column you intend to use.
-
Q.What kind of Sep-pak should be used for the assay?
-
A.We use Sep-pak light.
-
Q.Is a correlation data with #27165 available?
-
A.Yes, it is. The data is listed on the product detail page as a photo image. Please refer to the data.
-
Q.Can I use "reconstituted labeled antibody" after re-stored at -20℃?
-
A.Not recommended. Please use labeled antibody immediately after the reconstitution.
-
Q.How should I conduct Endothelin extraction process?
-
A.Please refer to the datasheet enclosed the product package and/or the datasheet listed on the product page.
As several publications can be found on the page (product description), please refer to them.
https://www.ibl-japan.co.jp/en/search/product/detail/id=3866 -
Q.Is extraction process essential for blood samples?
-
A.Yes, it is essential. Please follow the datasheet for the extraction process using column such as SepPak.
-
Q.I have only 0.5mL blood samples. Is it possible to measure it using this ELISA?
-
A.Yes, it is possible, however, in general speaking, the concentration of Endothelin in blood sample is low conc., the concentration could be much lower after the extraction and process. Please well consider the risk of lack of concentration prior start the measurement.
Our recommendation is more than 2mL blood samples.
-
Q.Can this kit be used for crab-eating macaque samples?
-
A.We do not have any data related to those types of samples.
-
Q.Can this kit be used for rat samples?
-
A.No, it cannot.
-
Q.Can this kit be used for bovine samples?
-
A.No, it cannot.
-
Q.What is the epitope of the capture antibody Anti- Osteopontin N-Half(34E3) Mouse IgG MoAb.?
-
A.It is SVVYGLR where is the position of thrombin cleavage site of human OPN so that this antibody specifically recognize the position at OPN N-Half.
-
Q.What kind of Osteopontin isoforms can be measured by this kit?
-
A.This kit can specifically measure human OPN N-Half that cleaved by thrombin.
-
Q.Can this kit be used for serum samples?
-
A.Osteopontin is rapidly digested by Thrombin in serum samples. For this reason, we do not recommend using serum samples for the assay because the value would be much lower.
-
Q.There are two types of EDTA-plasma tubes (Na and K). Which one is best to use?
-
A.Either tube can be used.
-
Q.What is the epitope of the pre-coated antibody (Anti-Human OPN (O-17) Rabbit IgG A.P.)?
-
A.The epitope is IPVKQADSGSSEEKQ.
-
Q.Can this kit be used to measure crab-eating macaque samples?
-
A.We do not have any data related to those types of samples.
-
Q.Can this kit be used for citrated-plasma samples?
-
A.We do not recommend using the kit for citrated-plasma samples.
-
Q.Can this kit be used for bovine samples?
-
A.No, it cannot.
-
Q.Is any sample stability data available?
-
A.Yes, it is available on the following relevant product page.
Please refer to the Pamphlet 2 listed on the page.
http://www.ibl-japan.co.jp/en/search/product/detail/id=3862 -
Q.What is the epitope of the detection antibody Anti-Human OPN(10A16)Mouse IgG MoAb Fab’A.P.?
-
A.It is KSKKFRRPDIQYPDATDE where is at the part of the right side position of thrombin cleavage site of human Osteopontin (OPN).
-
Q.What kind of Osteopontin isoforms can be measured by this kit?
-
A.This kit uses the antibody (#18625 OPN (O-17)) that recognizes the region at N-terminal of Osteopontin (OPN) as the capture antibody and and the antibody (#10011 OPN(10A16)) that recognizes the region at the part of the right side position of thrombin cleavage site of human OPN as the detection antibody. Therefore, this kit can detect all human OPN contains N-term region and the region described (10A16) above including Full Length of OPN. However, OPN cleaved by Thrombin cannot be measured by this kit. Please use #27258 Human Osteopontin N-Half Assay Kit - IBL for measurement of human OPN N-Half.
-
Q.Is there any cross reactivity other than VEGF161?
-
A.The specificity is described in “10. Kit performance” on the datasheet. Please refer to the datasheet for additional information.
-
Q.Can this kit be used for plasma samples?
-
A.EDTA-plasma can be used. However, heparin-plasma cannot be used for the assay.
-
Q.What is recombinant VEGF derived from?
-
A.It is a supernatant of CHO.
-
Q.Can this kit be used for mice and rat samples?
-
A.No, it cannot be used for those animal samples.
-
Q.What is the expression system of the standard?
-
A.It is a recombinant protein expression system. Wheat germ is used in this system.
-
Q.What is the recommended dilution ratio?
-
A.We cannot recommend a specific dilution ratio, but we generally use a dilution magnification of anywhere between 2 to 16 when we conduct measurements.
-
Q.What is the epitopes of the pre-coated antibodies used in this kit?
-
A.The epitope of the anntibody (Human LRG (162) Rabbit IgG A.P.) is GLKALGHLSGNRLRKL.
-
Q.What is the epitope of the labeled antibody?
-
A.The epitope of the labeled antibody (Human LRG (329) Rabbit IgG Fab’ A.P.) is AGPEAVKGQTLLAVAKSQ.
-
Q.Is there any data available for the measurement example and the reproducibility using normal controls?
-
A.Yes there is. The date is available on the following relevant product page. Please refer to the Pamphlet 2 listed on the page.
http://www.ibl-japan.co.jp/en/search/product/detail/id=3945 -
Q.Does this kit specifically detect LRG1 or other LRG families as well?
-
A.We have not actually checked that the cross reactivity against other LRG families using our LRG-1 kit #27769, however, as epitope of the antibodies that are used for ELISA kit are different positions from Leucine-rich repeat region, theoretically we believe that it specifically detects LRG1 and does not cross react to other LRG families.
-
Q.Can this kit be used for serum samples?
-
A.We do not recommend using serum samples for the kit.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL).
-
Q.Can this kit be used for the brain extracts from transgenic mice with Alzheimer's disease?
-
A.yes, it can.
-
Q.Can this kit be used for monkey samples?
-
A.Because the epitopes of monkeys and human beings are the same, we believe the kit can also be used for monkey samples. However, we do not have any data to back this claim.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL).
-
Q.Can this kit be used for brain extracts of Alzheimer's disease transgenic mice?
-
A.yes, it can.
-
Q.Can this kit be used for monkey samples?
-
A.Because the epitopes of monkeys and human beings are the same, we believe the kit can also be used for monkey samples. However, we do not have any data to back this claim.
-
Q.Is any special treatment required for serum samples?
-
A.No, no special treatment is required.
-
Q.Do you have any protocol for preparation of mice and/or rat brain
tissue extracts? -
A.CHAPS is recommended for extractin of wild type (W.T.) mice or rat brain tissue.
Please refer to the following link for the brain tissue extraction process using CHAPS.
http://www.ibl-japan.co.jp/files/topics/4944_ext_02_0.pdf
Guanidine solution is recommended for extraction of Transgenic (Tg) mice or rat brain tissue.
Please refer to the following link and the following two scientific articles for brain tissue extraction process using Guanidine.
http://www.ibl-japan.co.jp/files/topics/4945_ext_02_0.pdf
Presenilin-2 mutation causes early amyloid accumulation and memory impairment in a transgenic mouse model of Alzheimer's disease. Toda T et al. J Biomed Biotechnol. 2011;2011:617974.
https://www.ncbi.nlm.nih.gov/pubmed/?term=21234330
Methylene blue modulates β-secretase, reverses cerebral amyloidosis, and improves cognition in transgenic mice. Mori T et al. J Biol Chem. 2014 Oct 31;289(44):30303-17.
https://www.ncbi.nlm.nih.gov/pubmed/?term=25157105
* The ELISA kit #27712 Aβ42 and #27714 Aβ40 have been upgraded to #27719 Aβ (1-42)(FL) and #27718 Aβ (1-40)(FL).
-
Q.Can this kit be used for mouse or rat samples?
-
A.No, it cannot.
-
Q.What is the epitope of antibodies used in this kit?
-
A.Anti-Human Syndecan-4 Rabbit IgG (Precoated Antibody) recognizes position no. 45-57 amino acid of human Syndecan-4.
Anti-Human Syndecan-4 Rabbit IgG (HRP Antibody) recognizes position no. 19-145 amino acid of human Syndecan-4. -
Q.What is the immunogen of 2 antibodies used in the kit?
-
A.Synthesized Peptide was used as the immunogen for Anti-Human Syndecan-4 Rabbit IgG (Precoated Antibody).
Recombinant Protein was used as the immunogen for Anti-Human Syndecan-4 Rabbit IgG (HRP Antibody).
-
Q.What is the recommended dilution ratio for plasma samples?
-
A.We recommend 40 fold dilution.
-
Q.What is recommended dilution ratio for human blood samples?
-
A.It is 40 or more than 40 fold.
-
Q.Where the epitope (recognizing region) of antibodies used in this kit are?
-
A.The recognizing region of the precoated polyclonal antibody (Anti- (Pro)renin receptor Rabbit IgG Affinity Purify) is located at the amino acid numbers of 148 to 165 in the sequence of (pro)renin receptor.
The recognizing region of the HRP labeled monoclonal antibody (Anti- (P)RR (93A1B) Mouse IgG Fab’ Affinity Purify) is located at the amino acid numbers of 17 to 88 in the sequence of (pro)renin receptor.
-
Q.What is the epitope of the pre-coated antibody (Anti-Mouse APP(599) Rabbit IgG)?
-
A.The epitope is DAEFGHDSGFEVRHQK.
-
Q.Does this kit detect sAPP695α, sAPP751α and sAPP770α?
-
A.Yes it does.
As this kit is a sandwich ELISA that uses antibodies at N-Terminal and C-Terminal of sAPPα, it detects total sAPPα inclusive sAPP695α, sAPP751α and sAPP770α.
-
Q.Can this kit be used for supernatant samples that contain FBS?
-
A.Yes it can.
-
Q.Can this kit be used for cleaved type of Osteopontin?
-
A.No it cannot.
-
Q.Can this kit be used for serum samples?
-
A.No, it cannot.
-
Q.Can this kit be used for heparin-plasma samples?
-
A.We do not recommend using heparin plasma with the kit. The use of this plasma can affect the measurement results.
-
Q.Can this kit be used for bovine samples?
-
A.No, it cannot.
-
Q.What is the epitope of the pre-coated antibody (anti-human APP OX2 (351) Rabbit IgG A.P.)?
-
A.The epitope is not available because the rights belong to a third party.
-
Q.Can this kit be used for mouse or rat samples?
-
A.No, it cannot.
-
Q.What are the epitopes of the antibodies used in this kit?
-
A.The epitope of the pre-coated antibody (Anti-Human AGT (72) Rabbit IgG Affinity Purify) is IQAKTSPVDEKALQDQLVLV. We do not have the epitope for HRP conjugated Anti- Human AGT (601) Mouse IgG, but the detection part is around C-terminal.
-
Q.What is the best time for collecting urine?
-
A.We recommend collecting urine on an as needed basis.
-
Q.How should urine samples be stored after they are collected?
-
A.They should be stored at -80°C immediately after they are collected.
-
Q.How many minutes should the urine samples be centrifuged, and at what rpm rate?
-
A.This process is not required.
-
Q.Do you have any data of normal value of total Angiotentinogen in human blood and standard range?
-
A.Please refer to the following article. https://www.ncbi.nlm.nih.gov/pubmed/?term=17553939
-
Q.Can dog sample be used for this kit?
-
A.The epitope of human and dog are quite different. Therefore, the kit cannot be used for dog sample.
-
Q.Can this kit be used for mouse sample?
-
A.No, it cannot.
-
Q.Can this kit be used for dog samples?
-
A.No, it cannot.
-
Q.Can serum or plasma samples be used for this kit?
-
A.No, they cannot.
-
Q.Can this kit be used for hamster samples?
-
A.No, it cannot.
-
Q.Can heparin-plasma be used for this kit?
-
A.Not recommended to use heparin-plasma samples as cytokine is influenced by heparin, the data would not be stable.
-
Q.Can this kit be used for serum or plasma samples?
-
A.We do not have any data related to those types of samples.
-
Q.What is the epitope of the pre-coated antibody (anti-human Gal-3 (50A3) mouse IgG MoAb)?
-
A.The epitope of 50A3 is the 64 amino acid residue of the N-terminal region of Galectin-3.
-
Q.Does this antibody (the anti-human Gal-3 (50A3)) only bind to the 64th a.a. residue of Galectin-3?
-
A.The antibody reacts to a.a. parts 1 to 64, which means the antibody can detect any part of Galectin-3 found in this portion (1 to 64). Thus, the antibody can also detect the full length of Galectin-3.
-
Q.What is the standard derived from?
-
A.It is derived from the supernatant of the human nerve glioblastoma cells (U251).
-
Q.Can this kit be used for heparin-plasma samples?
-
A.We do not recommend using the kit for heparin plasma samples.
- 27101
- Rat VEGF Assay Kit - IBL
-
Q.Can this kit be used for serum samples?
-
A.No, it cannot.
-
Q.Can this kit be used for monkey samples?
-
A.The GIP sequences for human beings and monkeys are exactly same. Thus, theoretically our human GIP (both total and active) can be used to measure monkey GIP. However, we have never used our kits to evaluate monkey GIP, so we cannot be 100% certain. Thus, we cannot make any guarantee regarding the measurement results.
-
Q.Can this kit be used for monkey samples?
-
A.The GIP sequences for human beings and monkeys are exactly same. Thus, theoretically our human GIP (both total and active) can be used to measure monkey GIP. However, we have never used our kits to evaluate monkey GIP, so we cannot be 100% certain. Thus, we cannot make any guarantee regarding the measurement results.
- 27363
- Rat DMP1 Assay Kit - IBL
-
Q.Can this kit be used for mouse samples?
-
A.No, it cannot.
-
Q.What is the standard derived from?
-
A.It is derived from animal cells.
-
Q.What is overnight determination? How much time is required?
-
A.More than 16 hours is required.
-
Q.What is the immunogen?
-
A.The immunogen is recombinant the protein of ANGPTL2.
-
Q.Can this kit be used for mouse or rat samples?
-
A.No, they cannot.
-
Q.Can this kit be used for mouse or rat samples?
-
A.No, they cannot.